CVE-2021-37160

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) during a File Upload for a firmware update.
Configurations

Configuration 1


Information

Published : 2021-08-02 01:15

Updated : 2022-07-12 05:42


NVD link : CVE-2021-37160

Mitre link : CVE-2021-37160

Products Affected
No products.
CWE
CWE-347

Improper Verification of Cryptographic Signature